联系我们

与泰克代表实时聊天。 工作时间:上午 9:00 - 下午 5:00(太平洋标准时间)。

电话

致电我们

工作时间:上午9:00-下午5:00(太平洋标准时间)

下载

下载手册、产品技术资料、软件等:

下载类型
型号或关键字

反馈

常见问题

常见问题
How do you calculate the amount of time the Spectrum Time indicator (the orange bar) represents?
Spectrum Time = Window Factor / Resolution BandwidthWindow Factors: FFT WindowFactor Kaiser2.23 Rectangular0.89 Hamming1.30 Hanning1.44 Blackman-Harris1.90 Flat-Top3.77 The Resolution Bandwidth can be found next to the RBW readout on the MDO4000 display.

此常见问题适用于:
50016
After sending the CURVE? query why do I only get 5000 points even though the record length is more than 5000 points?
It is possible the DATA:STOP value is set to 5000 which is a common default value. When transferring data from the oscilloscope the user must specify a DATA:START and DATA:STOP value or the default values will be used.

此常见问题适用于:
52346
Are there IBIS or Spice loading models available for your Logic Analyzer probes?
Unfortunately we do not have pre-developed IBIS/Spice Models using the information provided in the P68xx probe manual.

此常见问题适用于:
52351
Are there programming examples of using TekVisa?
When TekVisa is installed a folder named “Visa Samples” is also installed in the C:\VXIpnp\WINNT\TekVISA\VisaSamples folder. There are also examples in the Tekvisa programmers manual located at: /dpo3012-manual/tekvisa An alternative to this sample …

此常见问题适用于:
52361
AWG5000: How do I enable the second monitor?
Minimize the AWG application then right click on an empty area of the screen. Select Properties then the Settings tab and click on monitor 2. Now check “extend my Windows desktop to this monitor” and press Apply.

此常见问题适用于:
52366
AWG5000: What is the SFDR on the AWG5000?
SFDR for the AWG5k series is shown in the list below.AWG5012 and AWG5014Normal Mode:50 dBc (37.5 MHz, 1.2 GS/s, 2.0 Vpk-pk)60 dBc (10 MHz, 600 MS/s, 1.0 Vpk-pk)80 dBc (1 MHz, 600 MS/s, 1.0 Vpk-pkDirect Mode:64 dBc (10 MHz, 600 MS/s, 0.6 Vpk-pk)80 dBc (1 …

此常见问题适用于:
52371
AWG7000: How do I enable the second monitor?
Minimize the AWG application then right click on an empty area of the screen. Select Properties then the Settings tab and click on monitor 2. Now check “extend my Windows desktop to this monitor” and press Apply.

此常见问题适用于:
52376
Can a Tektronix Encore product be reserved?
Reconditioned products are subject to availability upon receipt of a purchase order. A unit cannot be reserved without a purchase order.

此常见问题适用于:
52381
Can I create a file (reference memory) in an external application and have the Logic Analyzer use the compare function with acquired data to compare the differences between the two?
The TLA Application Software cannot compare against an external source. The target has to be another saved .tla file acquired through the TLA probe interface.

此常见问题适用于:
52386
Can my TLA600 be upgraded from Windows 2000 to Windows XP and still run version 4.4 of the TLA software?
The TLA600 is not supported under Windows XP. Windows 2000 and application software v4.4 was the last supported software configuration.

此常见问题适用于:
52391