联系我们

与泰克代表实时聊天。 工作时间:上午 9:00 - 下午 5:00(太平洋标准时间)。

电话

致电我们

工作时间:上午9:00-下午5:00(太平洋标准时间)

下载

下载手册、产品技术资料、软件等:

下载类型
型号或关键字

反馈

DPO4034

此产品已经停止销售,请与我们联系获得更多帮助。

Characteristics

Vertical System

Characteristic

DPO4034

DPO4054

DPO4104

Input Channels

4

4

4

Analog Bandwidth (-3 dB)

5 mV/div - 1 V/div

350 MHz

500 MHz

1 GHz

Calculated Rise Time

5 mV/div (typical)

1 ns

700 ps

350 ps

Hardware Bandwidth Limits

20 MHz or 250 MHz

Input Coupling

AC, DC, GND

Input Impedance

1 MΩ ±1%, 50 Ω ±1%

Input Sensitivity, 1 MΩ

1 mV/div to 10 V/div

Input Sensitivity, 50 Ω

1 mV/div to 1 V/div

Vertical Resolution

8 bits (11 bits with Hi Res)

Max Input Voltage, 1 MΩ

250 VRMS with peaks ≤ ±400 V

Max Input Voltage, 50 Ω

5 VRMS with peaks ≤ ±20 V

DC Gain Accuracy

±1.5% with offset set to 0 V

Offset Range

1 MΩ

50 Ω

   1 mV/div to 50 mV/div

±1 V

±1 V

   50.5 mV/div to 99.5 mV/div

±0.5 V

±0.5 V

   100 mV/div to 500 mV/div

±10 V

±10 V

   505 mV/div to 995 mV/div

±5 V

±5 V

   1 V/div to 5 V/div

±100 V

±5 V

   5.05 V/div to 10 V/div

±50 V

NA

Channel-to-Channel Isolation

≥100:1 at ≤100 MHz and ≥30:1 at >100 MHz up to the rated bandwidth for any two channels having equal volts/div settings

Horizontal System

Characteristic

DPO4034

DPO4054

DPO4104

Maximum Sample Rate (all channels)

2.5 GS/s

2.5 GS/s

5 GS/s

Maximum Record Length (all channels)

10 M points

Maximum Duration at Highest Sample Rate (all channels)

4 ms

4 ms

2 ms

Time Base Range (s/div)

1 ns to 400 s

400 ps to 400 s

Time Base Delay Time Range

-10 divisions to 50 s

Channel-to-Channel Deskew Range

±100 ns

Time Base Accuracy

±5 ppm over any ≥1 ms interval

Trigger System

Characteristic

Description

Main Trigger Modes

Auto, Normal, and Single

Trigger Coupling

DC, HF reject (attenuates >50 kHz), LF reject (attenuates

Trigger Holdoff Range

20 ns to 8 s

Sensitivity

   Internal DC Coupled

0.4 div DC to 50 MHz increasing to 1 div at rated bandwidth

   External

   (auxiliary input)

200 mV from DC to 50 MHz increasing to 500 mV at 250 MHz

Trigger Level Range

   Any Channel

±8 divisions from center of screen

   External

   (auxiliary input)

±8 V

Acquisition Modes

Mode

Description

Sample

Acquire sampled values

Peak Detect

Captures glitches as narrow as 200 ps at all sweep speeds

Averaging

From 2 to 512 waveforms included in average

Envelope

Min-max envelope reflecting Peak Detect data over multiple acquisitions

Hi Res

Real-time boxcar averaging reduces random noise and increases resolution

Roll

Scrolls waveforms right to left across screen at sweep speeds slower than or equal to 40 ms/div

Trigger Modes

Mode

Description

Edge

Positive or negative slope on any channel or front-panel auxiliary input. Coupling includes DC, HF reject, LF reject, and noise reject

Pulse Width

Trigger on width of positive or negative pulse that are >,

Runt

Trigger on a pulse that crosses one threshold but fails to cross a second threshold before crossing the first again

Logic

Trigger when any logical pattern of channels goes false or stays true for specified period of time. Any input can be used as a clock to look for the pattern on a clock edge. Pattern (AND, OR, NAND, NOR) specified for four input channels defined as High, Low, or Don't Care

Setup and Hold

Trigger on violations of both setup time and hold time between clock and data present on any two input channels

Rise/Fall Time

Trigger on pulse edge rates that are faster or slower than specified. Slope may be positive, negative, or either

Video

Trigger on all lines, odd, even, or all fields on NTSC, PAL, and SECAM video signals

Extended Video (optional)

Trigger on 480p/60, 576p/50, 720p/30, 720p/50, 720p/60, 875i/60, 1080i/50, 1080i/60, 1080p/24, 1080p/24sF, 1080p/25, 1080p/30, 1080p/50, 1080p/60, and custom bi-level and tri-level sync video standards

I2C (optional)

Trigger on Start, Repeated Start, Stop, Missing ACK, Address (7 or 10 bit), Data, or Address and Data on I2C buses up to 3.4 Mb/s

SPI (optional)

Trigger on SS, Idle Time, MOSI, MISO, or MOSI and MISO on SPI buses up to 10.0 Mb/s

CAN (optional)

Trigger on Start of Frame, Frame Type (data, remote, error, overload), Identifier (standard or extended), Data, Identifier and Data, End of Frame, or Missing ACK, or Bit Stuffing Errors on CAN signals up to 1 Mb/s. Data can be further specified to trigger on ≤, , ≥, or ≠ a specific data value. User-adjustable sample point is set to 50% by default

RS-232/422/485/UART (optional)

Trigger on Tx Start Bit, Rx Start Bit, Tx End of Packet, Rx End of Packet, Tx Data, and Rx Data

LIN (optional)

Trigger on Sync, Identifier, Data, Identifier and Data, Wakeup Frame, Sleep Frame, or Errors such as Sync, Parity, or Checksum Errors

FlexRay (optional)

Trigger on Start of Frame, Type of Frame (Normal, Payload, Null, Sync, Startup), Identifier, Cycle Count, Complete Header Field, Data, Identifier and Data, End of Frame or Errors such as Header CRC, Trailer CRC, Null Frame, Sync Frame, or Startup Frame Errors.

Trigger Delay by Time

4 ns to 8 s

Trigger Delay by Events

1 to 9,999,999 events

Waveform Measurements

Characteristic

Description

Cursors

Waveform and Screen

Automatic Measurements

29, of which up to eight can be displayed on screen at any one time. Measurements include Period, Frequency, Delay, Rise Time, Fall Time, Positive Duty Cycle, Negative Duty Cycle, Positive Pulse Width, Negative Pulse Width, Burst Width, Phase, Positive Overshoot, Negative Overshoot, Peak to Peak, Amplitude, High, Low, Max, Min, Mean, Cycle Mean, RMS, Cycle RMS, Positive Pulse Count, Negative Pulse Count, Rising Edge Count, Falling Edge Count, Area, and Cycle Area

Measurement Statistics

Mean, Min, Max, Standard Deviation

Reference Levels

User-definable reference levels for automatic measurements can be specified in either percent or units

Gating

Isolate the specific occurrence within an acquisition to take measurements, on using either the screen or waveform cursors

Waveform Math

Option

Description

Arithmetic

Add, subtract, multiply, and divide waveforms

Math Functions

Integrate, Differentiate, FFT

FFT

Spectral magnitude. Set FFT Vertical Scale to Linear RMS or dBV RMS, and FFT Window to Rectangular, Hamming, Hanning, or Blackman-Harris

Advanced Math

Define extensive algebraic expressions including waveforms, math functions, scalars, up to two user-adjustable variables, and results of parametric measurements e.g. (Intg(Ch1-Mean(Ch1)) × 1.414 × VAR1)

Software

Application

Description

NI LabVIEW SignalExpress™ Tektronix Edition LE

A fully interactive measurement software environment optimized for the DPO4000 Series, enables you to instantly acquire, generate, analyze, compare, import, and save measurement data and signals using an intuitive drag-and-drop user interface that does not require any programming. Standard DPO4000 Series support for acquiring, controlling, viewing, and exporting your live signal. The full version (SIGEXPTE) adds additional signal processing, advanced analysis, mixed signal, sweeping, limit testing, and user-defined step capabilities and is available for a 30-day trial period standard with each instrument

OpenChoice® Desktop

Enables fast and easy communication between a Windows PC and the DPO4000 Series, through USB or LAN. Transfer and save settings, waveforms, measurements, and screen images.

IVI Driver

Provides a standard instrument programming interface for common applications such as LabVIEW, LabWindows/CVI, Microsoft .NET, and MATLAB

Display Characteristics

Characteristic

Description

Display Type

10.4 in. (264 mm) liquid-crystal TFT color display

Display Resolution

1,024 (H) × 768 (V) pixels (XGA)

Waveform Styles

Vectors, Dots, Variable Persistence, Infinite Persistence

Graticules

Full, Grid, Cross Hair, Frame, IRE, and mV

Format

YT and simultaneous XY/YT

Waveform Capture Rate

Up to 50,000 wfms/s

Input/Output Ports

Port

Description

CompactFlash Drive

Front-panel access (Type 1)

USB 2.0 Full-speed Host Port

Supports USB mass storage devices and printers. Two ports available on rear panel and one on front panel

USB 2.0 High-speed Device Port

Rear-panel connector allows for control of oscilloscope through USBTMC or GPIB with a TEK-USB-488

LAN Port

RJ-45 connector, supports 10/100Base-T

XGA Video Port

DB-15 female connector, connect to show the oscilloscope display on an external monitor or projector

Auxiliary Input

Front-panel BNC connector. Input Impedance 1 MΩ. Max input 250 VRMS with peaks ≤ ±400 V

Probe Compensator Output

Front-panel pins. Amplitude 2.5 V. Frequency 1 kHz

Trigger Out

Rear-panel BNC connector, provides a positive polarity pulse when the oscilloscope triggers

Kensington Lock

Rear-panel security slot connects to standard Kensington lock

Power Source

Characteristic

Description

Power Source Voltage

100 to 240 V ±10%

Power Source Frequency

47 – 66 Hz (90 to 264 V)

360 – 440 Hz (100 to 132 V)

Power Consumption

250 W maximum

General Characteristics

Physical Characteristics

Dimensions

mm

in.

Height

229

9.0

Width

439

17.3

Depth

137

5.4

Weight

kg

lb.

Net

5

11

Shipping

10

22

Rackmount Configuration

5U

Cooling Clearance

2 in. (51 mm) required on left side and rear of instrument

Environmental

Characteristic

Description

Temperature

   Operating

0 °C to +50 °C

   Nonoperating

-20 °C to +60 °C

Humidity

   Operating

High: 40 °C to 50 °C, 10% to 60% Relative Humidity

Low: 0 °C to 40 °C, 10% to 90% Relative Humidity

   Nonoperating

High: 40 °C to 60 °C, 5% to 60% Relative Humidity

Low: 0 °C to 40 °C, 5% to 90% Relative Humidity

Altitude

   Operating

3,000 meters (9,843 feet)

   Nonoperating

12,000 meters (39,370 feet)

Random Vibration

   Operating

0.31 GRMS from 5 to 500 Hz, 10 minutes each axis, 3 axes, 30 minutes total

   Nonoperating

2.46 GRMS from 5 to 500 Hz, 10 minutes each axis, 3 axes, 30 minutes total

Regulatory

Electromagnetic Compatibility

89/336/EEC

Safety

UL61010-1, Second Edition; CSA61010-1 Second Edition, EN61010-1: 2001; IEC 61010-1: 2001