Specifications
All specifications are guaranteed unless noted otherwise. All specifications apply to all models unless noted otherwise.
Model overview
TBS1064 | TBS1104 | TBS1154 | |
---|---|---|---|
Bandwidth 1 | 60 MHz | 100 MHz | 150 MHz |
Channels | 4 | 4 | 4 |
Sample rate on each channel | 1.0 GS/s | 1.0 GS/s | 1.0 GS/s |
Record length | 2.5k points at all time bases |
1Bandwidth is 20 MHz at 2 mV/div
Vertical system — Analog channels
- Vertical resolution
- 8 bits
- Input sensitivity range
- 2 mV to 5 V/div on all models with calibrated fine adjustment
- DC gain accuracy
- ±3%, from 10 mV/div to 5 V/div
- Maximum input voltage
- 300 VRMS CAT II; derated at 20 dB/decade above 100 kHz to 13 Vp-p AC at 3 MHz
- Offset range
-
2 mV to 200 mV/div: ±1.8 V
>200 mV to 5 V/div: ±45 V
- Bandwidth limit
- 20 MHz
- Input coupling
- AC, DC, GND
- Input impedance
- 1 MΩ in parallel with 20 pF
- Vertical zoom
- Vertically expand or compress a live or stopped waveform
Horizontal system — Analog channels
- Time base range
- 5 ns to 50 s/div
- Time base accuracy
- 50 ppm
- Horizontal zoom
- Horizontally expand or compress a live or stopped waveform
Input/Output ports
- USB interface
- USB host port on front panel supports USB flash drives
- GPIB interface
- Optional
Data storage
- Nonvolatile storage
-
- Reference waveform display
- 2.5K point reference waveforms
- Waveform storage without USB flash drive
- 2.5K point
- Maximum USB flash drive size
- 64 GB
- Waveform storage with USB flash drive
- 96 or more reference waveforms per 8 MB
- Setups without USB flash drive
- 10 front-panel setups
- Setups with USB flash drive
- 4000 or more front-panel setups per 8 MB
- Screen images with USB flash drive
- 128 or more screen images per 8 MB (the number of images depends on file format selected)
- Save All with USB flash drive
- 12 or more Save All operations per 8 MB
A single Save All operation creates 3 to 9 files (setup, image, plus one file for each displayed waveform)
Acquisition system
- Acquisition modes
-
- Peak Detect
- High-frequency and random glitch capture. Captures glitches as narrow as 12 ns (typical) at all time base settings from 5 μs/div to 50 s/div
- Sample
- Sample data only
- Average
- Waveform averaged, selectable: 4, 16, 64, 128
- Single Sequence
- Use the Single Sequence button to capture a single triggered acquisition sequence
- Roll
- At acquisition time base settings of >100 ms/div
Trigger system
- External trigger input
- Included on all models
- Trigger modes
- Auto, Normal, Single Sequence
- Trigger types
-
- Edge (Rising/Falling)
- Conventional level-driven trigger. Positive or negative slope on any channel. Coupling selections: AC, DC, Noise Reject, HF Reject, LF Reject
- Video
- Trigger on all lines or individual lines, odd/even or all fields from composite video, or broadcast standards (NTSC, PAL, SECAM)
- Pulse Width (or Glitch)
- Trigger on a pulse width less than, greater than, equal to, or not equal to, a selectable time limit ranging from 33 ns to 10 s
- Trigger source
- Four channel models: CH1, CH2, CH3, CH4, Ext, Ext/5, AC Line
- Trigger view
- Displays trigger signal while Trigger View button is depressed.
- Trigger signal frequency readout
- Provides a frequency readout of the trigger source.
Waveform measurements
- Cursors
-
- Types
- Amplitude, Time
- Measurements
- ΔT, 1/ΔT, ΔV
- Automatic measurements
- Period, Frequency, +Width, –Width, Rise Time, Fall Time, Max, Min, Peak-to-Peak, Mean, RMS, Cycle RMS, Cursor RMS, Duty Cycle, Phase, and Delay
Waveform math
- Arithmetic
- Add, Subtract, Multiply
- Math functions
- FFT
- FFT
- Windows: Hanning, Flat Top, Rectangular
- Sources
- CH1 – CH2, CH2 – CH1, CH1 + CH2, CH1 × CH2, CH3 – CH4, CH4 – CH3, CH3 + CH4, CH3 × CH4
Autoset
- Autoset menu
- Single-button, automatic setup of all channels for vertical, horizontal, and trigger systems, with undo Autoset
- Square wave
- Single Cycle, Multicycle, Rising or Falling Edge
- Sine wave
- Single Cycle, Multicycle, FFT Spectrum
- Video (NTSC, PAL, SECAM)
- Field: All, Odd, or Even
Autorange
- Automatically adjust vertical and/or horizontal oscilloscope settings when probe is moved from point to point, or when the signal exhibits large changes.
Display system
- Interpolation
- Sin (x)/x
- Waveform styles
- Dots, vectors
- Persistence
- Off, 1 s, 2 s, 5 s, infinite
- Format
- YT and XY
Physical characteristics
- Dimensions
-
- Height
- 158.0 mm (6.22 in)
- Width
- 326.3 mm (12.85 in)
- Depth
- 124.2 in (4.89 in)
- Shipping dimensions
-
- Height
- 266.7 mm (10.5 in)
- Width
- 476.2 mm (18.75 in)
- Depth
- 228.6 mm (9.0 in)
- Weight
-
- Instrument only
- 2.0 kg (4.4 lbs)
- ... with accessories
- 2.2 kg (4.9 lbs)
- RM2000B rackmount
-
- Width
- 482.6 mm (19.0 in)
- Height
- 177.8 mm (7.0 in)
- Depth
- 108.0 mm (4.25 in)
Environmental
- Temperature
-
- Operating
- 0 to +50 ºC
- Nonoperating
- –40 to +71 ºC
- Humidity
-
- Operating and nonoperating
-
Up to 85% RH at or below +40 ºC
Up to 45% RH up to +50 ºC
- Altitude
-
- Operating and nonoperating
- Up to 3,000 m (9,843 ft.)
- Regulatory
-
- Electromagnetic compatibility
- Meets Directive 2004/108/EC, EN 61326-2-1 Class A; Australian EMC Framework
- Safety
- UL61010-1:2004, CSA22.2 No. 61010-1:2004, EN61010-1:2001, IEC61010-1:2001