Characteristics

Signal Acquisition System

Bandwidth -

TDS210: 60 MHz

TDS220: 100 MHz

TDS240: 100 MHz

Sample Rate - 1 GS/s on each channel

Channels -

TDS210 and TDS220: 2 identical channels plus external trigger

TDS224: 4 identical channels

Sensitivity (with calibrated fine adjustment) - 2 mV to 5 V/div (bandwidth limited to 20 MHz at 2 mV/div and 5 mV/div, in all modes, and 20 MHz at 10 mV/div in Peak Detect mode)

Vertical Zoom - Vertically expand or compress a live or stopped waveform

Calibrated Position Range

Volts/Div Setting

Offset Range

2 mV to 200 mV/div

±2 V

>200 mV to 5 V/div

±50 V

DC Gain Accuracy - ±3% (±4% at 2 mV/div and 5 mV/div)

Vertical Resolution - 8-bit (256 levels over 10.24 vertical divisions)

Automatic Measurements

Period, frequency, cycle RMS, mean, peak-to-peak

Acquisition Modes

Sample, Average, Peak Detect - High frequency and random glitch capture; captures glitches as narrow as 10 ns using acquisition hardware at all time/div settings between 5 µs/div and 5 s/div

Time Base System (main and Window)

Horizontal Zoom - Horizontally expand or compress a live or stopped waveform

Time/Division Range - 5 ns to 5 s/div

Record Length - 2,500 sample points per channel

Horizontal Accuracy - ±0.01%

Nonvolatile Storage

Waveform Display - Two 2,500 point reference waveforms

Waveform Storage -

TDS210 and TDS220: Two 2,500 point reference waveforms

TDS224: Four 2,500 point reference waveforms

Setups - 5 front panel setups

Triggering System (main Only)

Trigger Type - Edge (rising or falling), video, set to 50%

Video Trigger Type - Triggers on fields or lines from composite video; triggers on broadcast standard NTSC, PAL or SECAM video

Trigger Modes - Auto, normal, single sweep

Trigger Source -

TDS210 and TDS220: CH1, CH2, Ext, Ext/5 and AC Line

TDS224: CH1, CH2, CH3, CH4 and AC Line

Trigger View - Displays trigger signal while trigger view button depressed.

Cursors

Types - Voltage, time

Measurements - Delta T, 1/Delta T, Delta V

Waveform Processing

Arithmetic Operators - Add, subtract

Sources - CH1±CH2, CH3±CH4 (TDS224)

Autoset - Single button automatic setup on selected input signal for vertical, horizontal and trigger systems.

Display System

Robust, backlit LCD with adjustable multilevel contrast.

Interpolation - Sin(x)/x

Display Modes - Vectors, dots

Persistence - Off, 1 sec, 2 sec, 5 sec, infinite

Format - YT and XY

Hardcopy Capability (requires Modules)

Printer/File Formats - Thinkjet, Deskjet, Laserjet, Epson (9- or 24-Pin), BMP, PCX, IMG, EPS, DPU 411, DPU 412

Hardcopy Layout - Landscape or portrait

TDS2CM Communications Extension Module

Centronics-type Parallel Port

RS-232 Programmability - Full talk/listen modes. Control of all modes, settings and measurements. Baud rate up to 19,200. 9-Pin, DTE.

GPIB Programmability - Full talk/listen modes. Control of all modes, settings and measurements (IEEE Std 488-1987). Includes WaveStar™ Software for Oscilloscopes 30 day product demo.

TDS2MM Measurement Extension Module

FFT -

Windows: Hanning, Flat Top, Rectangular

Sample points: 2,048

Automatic Measurements - Rise/fall time, positive/negative pulse width

Interface - Centronics, RS-232, GPIB

Environmental and Safety

Temperature -

Operating: 0°C to +50°C

Nonoperating: -20°C to +60°C

Humidity -

Up to 90% RH at or below +40°C;

Up to 60% RH from 41°C to 50°C (operating and nonoperating)

Altitude - Up to 3,000 m (operating)

EMC - Meets Directive 89/336/EEC, GOST

Safety - UL3111-1, CSA1010.1, IEC61010-1, EN61010-1

Physical Characteristics

Dimensions

mm

in.

Weight

kg

lb.

Width

304.8

12.0

Height

151.4

5.96

Depth

120.7

4.75

Oscilloscope only

1.5

3.25

w/accessories

1.7

3.75