联系我们

与泰克代表实时聊天。 工作时间:上午 9:00 - 下午 5:00(太平洋标准时间)。

电话

致电我们

工作时间:上午9:00-下午5:00(太平洋标准时间)

下载

下载手册、产品技术资料、软件等:

下载类型
型号或关键字

反馈

Why is the TDS7254/B not recommended for more than XGA resolution?

问题:

Why is the TDS7254/B not recommended for more than XGA resolution?

答案:

TDS7254/B Oscilloscopes typical rise time is 83ps for 20% to 80% signal rise time. As a rule of thumb, the Oscilloscope rise time should be at least 3 to 4 times faster than the signal rise time. For XGA resolution the compliance test specification rise time limit is around 400ps, but the designs will be around 200ps to 300ps. This is with in the range of TDS7254/B Oscilloscope. For the higher resolutions the signals rise time will be less than 200ps, so it is not recommended to use TDS7254/B Oscilloscopes.

此常见问题适用于:

无产品系列

产品:

常见问题 ID 60296

查看所有常见问题 »