Single-Ended probes are mentioned for HDMI intra-pair skew measurements. Can we use the probe P7330 for single ended measurements?

Subject: Compatibility | Published: October 2026 | FAQ ID: KB-02877

Question: Single-Ended probes are mentioned for HDMI intra-pair skew measurements. Can we use the probe P7330 for single ended measurements?

Answer: The differential input range of P7330 is ± 2.5V, but the typical HDMI signal voltage range is 2.9V to 3.3V, so it is not recommended to use the P7330 for Single ended measurements.

NOTE: this applies to testing under CTS1.0,1.1,1,2 only.