TDP1500 voltage range

Model: TDP1500 | Subject: Products | Published: October 2026 | FAQ ID: KB-02670

Question: TDP1500 voltage range

Answer: TDP1500 differential input dynamic range is:

±8.5 V ; 6 V RMS (10X)

±850 mV ; 0.6 V RMS (1X)

Max input voltage (non destructive) is ±25 V (DC + pk AC). For more specific details please refer to the Technical Reference document here https://www.tek.com/en/manual/low-voltage-differential-probe/tdp1500-differential-probe-low-voltage"