Thermocouple Datalogging and CJC option 2001-SCAN

Model: 2001-SCAN | Subject: Setup & Operation | Published: October 2026 | FAQ ID: KB-01060

Question: I don't see any means to set Chan#1 to CJC external on the 2001-SCAN?  I can place Ch2-9 into thermocouple Type T but don't have ability to set CJC or know if that's working correct for cold junction compensation.

Answer: Thank you for contacting Tektronix support.
Channel 1 on the 2001-SCAN card must be a reference junction, which is why only channel 2-9 are able to be set to thermocouple. https://www.tek.com/en/default-accessory-series-manual-77