What do the red/green lights on the RSA306 indicate?

Subject: Setup & Operation | Published: October 2026 | FAQ ID: KB-01888

Question: What do the red/green lights on the RSA306 indicate?

Answer:

Solid Green:  Instrument has finished boot-up up and is ready to connect or operate.

Red: Unable to power up, or re-booting the instrument

Flashing green: Transferring data

To troubleshoot connection issues, first make sure that the Tektronix USB driver is installed. To do this:

Open Device Manager on the PC.  Then look in "Universal Serial Bus Controllers".  You should see your RSAXXX instrument listed like "Tektronix RSA306B".  

If  you do not see your Tektronix RSAXXX listed – you will need to update the USB3.0 drivers from the motherboard manufacturers website.